NSN 6515-01-671-4127, 6515016714127
CONCENTRATOR,OXYGEN

Federal Supply Classification
FSC 6515 - Medical and Surgical Instruments, Equipment, and Supplies
National Item Identification Number
NIIN 016714127
Codification Country
United States
Item Name Code
INC 46708
Criticality
(X) This item does not have a nuclear hardened feature or any other critical feature such as tolerance, fit restriction or application.
Hazardous Material Indicator Code
(N) There is no data in the HMIRS and the NSN is in an FSC not generally suspected of containing hazardous materials.
Availability: Request Qualified Suppliers: Currently qualifying suppliers

Manufacture or supply this part?
Join our growing community of certified suppliers.

Let Us Know
  CALL ABOUT NSN 6515-01-671-4127
Our client representatives are standing by to help fulfill your request.

+1-321-773-5710

  REQUEST QUOTE for 6515-01-671-4127


Have a list that needs to be quoted? Upload it here.

DLMS® Summary

National Stock Number 6515-01-671-4127

National Stock Number (NSN) 6515-01-671-4127, or NIIN 016714127, (concentrator,oxygen) was assigned March 27, 2018 in the Federal Logistics Information System (FLIS). This NIIN is Available for Manual Assignment. This NSN does not replace any other NSNs.

There are 4 manufacturer part numbers associated with this NSN. None of the associated part number(s) are considered obsolete. The manufacturer part numbers were originally supplied by one supplier. Today, one supplier is listed as an Active supplier for this NSN. Of the active suppliers, only one supplier is deemed design controlled or valid supply parts.

This part number has not been procured by the US Government in over 5 years.

There has been a small amount of demand for this national stock number. The demand for this NSN originated from less than 5 countries. Based on a planned procurement review by the Primary Inventory Control Activity (PICA) on Mar 28, 2018, this NSN's acquisition method was noted as: Not established. Not established.

This NSN is assigned to Item Name Code (INC) 46708. [A PORTABLE INSTRUMENT CONSISTING OF A CALIBRATED ILLUMINATED FLOWMETER, BUILT-IN ALARMS FOR POWER FAILURE AND HIGH/LOW PRESSURE, FILTERS TO REMOVE DUST, BACTERIA, AND OTHER PARTICULATES, AND ALSO PROVIDES FOR THE ATTACHMENT OF AN IN-LINE HUMIDIFIER TO HUMIDIFY GAS MIXTURE DELIVERED TO THE PATIENT. IT IS DESIGNED FOR USE IN PRODUCING OXYGEN-RICH GAS MIXTURE BY DRAWING IN ROOM AIR AND EXTRACTING OXYGEN.]. This item does not have a nuclear hardened feature or any other critical feature such as tolerance, fit restriction or application. There is no data in the HMIRS and the NSN is in an FSC not generally suspected of containing hazardous materials. NSN 6515016714127 does not contain precious metals.

This NSN is associated to Schedule B 9018908000: other instruments and appliances used in medical, surgical, dental or veterinary sciences, nesoi. The Schedule B End Use is listed as medicinal equipment. NAICS classification category 339112: surgical and medical instrument manufacturing.

This information was last updated on .


Technical Characteristics

III PART NAME ASSIGNED BY CONTROLLING AGENCY
(CXCY) EXPEDITIONARY DEPLOYABLE OXYGEN CONCENTRATION SYSTEM 120 LPM
CURRENT TYPE
(ACDC) AC
FREQUENCY IN HERTZ
(FREQ) 50.0 OR 60.0
PART NAME ASSIGNED BY CONTROLLING AGENCY
(CXCY) EXPEDITIONARY DEPLOYABLE OXYGEN CONCENTRATION SYSTEM 120 LPM
SPECIAL FEATURES
(FEAT) PART NUMBER: 793980-001 EDOCS IS COMPLETELY CONTAINED IN A RUGGED ALUMINUM STRUCTURAL FRAME. OXYGEN PURITY IN ACCORDANCE WITH USP 93% PRODUCTION: 120 LITERS PER MINUTE MINIMUM FLOW AT USP 93% PRESSURE IS REGULATED TO 80 PSI DELIVERY PRESSURE TWO LOW-PRESSURE SUPPLY CONNECTIONS TWO HIGH-PRESSURE COMPRESSORS PROVIDES 2,250 PSI AT 120 LPM. FOUR "M" SIZED CYLINDERS (3,540 LITERS OF OXYGEN AT 2,250 PSI PER CYLINDER) MOUNTED WITHIN THE EDOCS FRAME PROVIDES BACK-UP STORAGE CAPACITY. FOUR "M" SIZED CYLINDERS EQUAL 14,160 LITERS OF OXYGEN. AT THE FULL-RATED FLOW OF 120 LITERS PER MINUTE, STORAGE WILL PROVIDE 2.0 HOURS OF BACKUP CAPABILITY. 208/240VAC 50/60HZ CYLINDER FILLING OXYGEN FOR FILLING CYLINDERS IS AVAILABLE FROM AN EDOCS AT A MAXIMUM RATE OF 120 LPM AT A PRESSURE OF 2,250 PSI. A VACUUM PUMP USED TO EVACUATE THE CYLINDER FROM ATMOSPHERIC PRESSURE TO 27 IN. HG OF VACUUM IS PROVIDED SO THAT ANY POSSIBLE CONTAINMENTS ARE REMOVED PRIOR TO REFILLING WITH OXYGEN.
VOLTAGE IN VOLTS
(ELEC) 208.0 AND 240.0